Features:
1.Advanced scanning technology with excellent SNR
2.Cutting-edge detection technology and most stable signal
3.Automatic gain function, accurate and reliable data
4.Exquisite and compact exterior with strong and durable structure
5.Touch operation interface, easy to operate
6.Complete function configuration and stable equipment performance
- parameter
- download
Model | FQD-16B | |||||||
Sample capacity | 16×0.2ml (suitable for single tube, 8-strips tubes, the transparent tube body and flat cover are required) | |||||||
Detection channel | F1 | F2 | F3 | F4 | ||||
Applicable dye | FAM, SYBR Green I | VIC, HEX, TET, JOE,
Cy3,
|
ROX
|
Cy5
|
||||
Temperature range of block working | 0~100℃ (Minimum division:0.1℃) | |||||||
Heating/cooling rate | ≥5℃/s | |||||||
Temperature fluctuation | ≤0.1°C (full-range), (55°C typical value ≤0.1°C) | |||||||
Temperature accuracy | ≤±0.1°C (full-range), (55°C typical value ≤0.1°C) | |||||||
Temperature uniformity | ≤0.25°C (full-range), (55°C typical value ≤0.15°C) | |||||||
Temperature range of hot-lid working | 70~110°C (adjustable, default 105°C) | |||||||
Repeatability of fluorescent intensity detection | 5% | |||||||
Running mode | Continuous running | |||||||
Operation system | Windows8/Windows10 | |||||||
Power supply | DC24V 180W | |||||||
Dimensions | 240mm×180mm×160mm | |||||||
Weight | 4kg |